Research Article | DOI: https://doi.org/10.31579/2767-7370/139
Department of Materials Engineering, Faculty of Engineering, Malayer University, Malayer, Iran.
*Corresponding Author: Dr. Ali Shanaghi and Dr. Ali Reza Souri, Department of Materials Engineering, Faculty of Engineering, Malayer University, Malayer, Iran.
Citation: Ali Shanaghi, Alireza Souri, Seyed Peyman Aiyoubian Markazi, Sajjad Lak, (2025), Electrochemically Deposited Biomimetic Hydroxyapatite-Collagen Coating on Ti6Al4V: Enhancing Corrosion Resistance and Nanomechanical Performance for Medical Applications, J New Medical Innovations and Research, 6(3); DOI:10.31579/2767-7370/139
Copyright: © 2025, Ali Shanaghi, Dr. Ali Reza Souri This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Received: 18 January 2025 | Accepted: 12 February 2025 | Published: 10 March 2025
Keywords: electrochemical deposition; hydroxyapatite-collagen; corrosion behavior; nano-mechanical properties; ti6al4v
This study investigated the synergistic effects of incorporating collagen type I into hydroxyapatite (HA) coatings on anodized Ti-6Al-4V substrates to enhance their biocompatibility and performance for biomedical applications. HA coatings were electrochemically deposited onto anodized Ti-6Al-4V, and collagen was subsequently incorporated via a soaking method. The coatings were characterized using GI-XRD, FESEM, AFM, nanoindentation, nanoscratch testing, and EIS. GI-XRD confirmed the presence of both anatase and rutile TiO2 in the anodized layer and revealed that collagen influenced HA crystallization, promoting the formation of aligned, elongated HA crystals. FESEM and AFM analyses showed a 30% reduction in surface roughness upon collagen incorporation, attributed to collagen filling HA pores. Nanoindentation testing demonstrated a decrease in Young's modulus from 0.2 GPa for HA to 0.1 GPa for the HA-collagen composite, and hardness decreased from 0.07 GPa to 0.01 GPa, respectively. Nanoscratch testing revealed improved wear resistance for the HA-collagen composite, with the coefficient of friction decreasing from 1.12 to 0.94 during scratching. EIS studies showed a two-fold increase in charge transfer resistance for the HA-collagen coating, indicating enhanced corrosion protection. These results suggest that the HA-collagen composite coating offers a promising approach for improving the long-term performance and clinical success of biomedical implants by enhancing cell adhesion, mitigating stress shielding, improving wear resistance, and enhancing corrosion protection.
Titanium (Ti) and its alloys are widely used as implant materials for replacing damaged load-bearing bones due to their excellent corrosion resistance, high mechanical strength, low density, and good biocompatibility [1]. Despite these advantages, titanium and its alloys face certain challenges when used as implants. Their bioinert nature often hinders bone tissue integration, and they may release undesirable ions, which can lead to adverse biological responses [2]. Additionally, titanium alloys are susceptible to reactions with corrosive ions present in body fluids, which can result in damage to adjacent tissues and structures [1–3]. Other issues, such as severe adhesive wear and stress shielding effects, further limit their performance as long-term implant materials [1–3]. To address these shortcomings, surface modifications, such as the hierarchical deposition of hydroxyapatite-collagen (Col/HA) composites on anodized Ti6Al4V alloys, have been proposed. These modifications aim to enhance biological compatibility and reduce the adverse effects associated with titanium-based implants [4].
Hydroxyapatite (HA), a key mineral component of human bone and teeth, is a calcium phosphate compound with the chemical formula Ca10(PO4)6(OH)2 [5–7]. Due to its excellent biocompatibility and osteoconductivity, HA-based nanocomposites have been extensively used in biomedical applications, including bone repair and regeneration [7–9]. Similarly, type I collagen, which constitutes about 90% of bone tissue, provides structural integrity and support for various connective tissues, including bone, skin, and tendons [10]. Recently, collagen has gained significant attention for its applications in tissue engineering and regenerative medicine [11, 12]. Combining HA with collagen in a nanocomposite coating offers the advantages of both materials, resulting in improved biocompatibility and osteoconductivity, along with the ability to mimic the hierarchical structure of natural bone [12–14].
Various coating techniques are available for producing HA layers on metallic substrates, including sol-gel [15–18], thermal and plasma spraying [19–22], biomimetic processes [23, 24], chemical vapor deposition [25], pulsed laser deposition [26, 27], and electrochemical deposition [28, 29]. Electrochemical deposition is an effective and versatile method for synthesizing HA coatings on Ti6Al4V alloys. This technique utilizes an applied electric field to deposit HA layers from an electrolyte solution onto the metallic substrate, offering advantages such as simplicity, low cost, and precise control over coating morphology and thickness [46]. Additionally, this method allows the formation of porous coatings, which facilitate bone ingrowth and improve implant fixation [43, 44]. However, the deposition process requires subsequent sintering at elevated temperatures to enhance the mechanical integrity and adhesion of the coatings. This sintering step, particularly at high temperatures exceeding 950 °C, can result in the decomposition of HA into undesirable calcium phosphate phases [50].
To ensure the integrity of nano-composite HA layers, sintering must be conducted at significantly lower temperatures compared to traditional microparticulate counterparts. These nano-composite coatings are derived from stable, non-agglomerated suspensions prepared using nano-sized starting powders. The use of a dispersing agent is critical to prevent particle agglomeration during the synthesis process. However, challenges arise during the sintering process, including the formation of cracks in the HA layer due to mismatched thermal expansion coefficients between the HA coating and the metallic substrate. Furthermore, significant differences in the chemical and structural properties of the materials hinder the formation of a strong chemical bond between the deposited HA layer and the Ti alloy substrate at their interface [50-59].However, this article introduces a novel approach to enhancing the performance of anodized Ti6Al4V implants by developing a hierarchical collagen-hydroxyapatite (Col/HA) nanocomposite coating, integrating advanced electrochemical and precipitation techniques. The research aims to overcome the limitations of traditional titanium implants, such as bioinertness and ion release, by mimicking the natural structure of bone tissue. The use of collagen, a key component of type I human bone, combined with hydroxyapatite—a primary constituent of bone minerals—results in a biomimetic coating that significantly improves corrosion resistance, mechanical properties, and biocompatibility. By employing optimized deposition techniques, this article demonstrates a novel strategy for achieving strong adhesion between the coating and Ti6Al4V substrate, along with enhanced nanomechanical performance and superior cellular response. This innovative work provides a significant leap toward the design of next-generation medical implants, offering improved durability, bioactivity, and compatibility for dental and orthopedic applications.
2.1. Substrat preparation
The experimental process commenced with the preparation of samples cut from a Grade 5 titanium sheet, commonly known as Ti-6Al-4V. Each sample was precisely machined to dimensions of 10 mm × 20 mm × 1 mm. To address any surface imperfections resulting from the cutting process, a systematic surface preparation protocol was employed. This involved sequential sanding using sandpaper with grit sizes of 800, 1000, 1500, 2000, 2500, and 3000. Following sanding, the samples were polished using a metallographic cloth in conjunction with an alumina suspension to achieve a high-quality finish. To ensure the removal of contaminants, the polished samples underwent a degreasing step. Each sample was immersed in acetone and then in 96% ethanol for a duration of 10 minutes. After degreasing, the samples were thoroughly rinsed with distilled water and allowed to air dry completely.
The anodization of the Ti6Al4V samples was performed in a sulfuric acid solution with a concentration of 10-15% at a controlled temperature of 25°C. The anodization process was carried out at a low voltage of 20 volts for a duration of 2 hours. This treatment resulted in the formation of a protective oxide layer on the titanium surface, enhancing its corrosion resistance and mechanical properties.
2.2. Coatings preparation
The hydroxyapatite (HA) coating solution was prepared by dissolving specific reagents in 100 mL of distilled water at a controlled temperature of 45°C. The components utilized for the solution preparation included: Calcium nitrate tetrahydrate, Ammonium dihydrogen phosphate and Sodium nitrate. The pH of the solution was carefully adjusted to approximately 4.5 using appropriate buffering agents. Following this adjustment, 1 mL of hydrogen peroxide was added to facilitate the reaction. The resulting solution was then placed in an ultrasonic bath at a temperature of 30°C for one hour to ensure complete homogeneity and uniform particle distribution. Then, the HA coating was applied via an electrochemical deposition process using a two-electrode configuration. In this setup, the titanium substrate served as the cathode while a stainless steel electrode acted as the anode. The substrates were immersed in the hydroxyapatite-titania solution and subjected to an electric potential of 3.1 volts at a temperature of 37°C for a duration of 15 minutes to promote effective coating formation.
Finally, the collagen coating was applied by dip coating method onto the hydroxyapatite-coated surface. Edible collagen powder was sourced from a pharmacy, and a solution was prepared by dissolving 500 mg of collagen in one liter of simulated body fluid (SBF) at room temperature. The HA-coated substrates were immersed in this collagen solution for a period of 24 hours to allow for effective adsorption and interaction between the collagen and HA layers. After immersion, the substrates were removed from the solution and allowed to air dry. This immersion-drying cycle was repeated three times to ensure adequate collagen layer formation. Finally, the coated samples were placed in an oven at a temperature of 80°C for 30 minutes to achieve thorough drying and stabilization of the coatings.
2.3. Coating characterizations
The characterization of HA and HA- collagen nanostructured coatings on anodized Ti6Al4V samples was conducted using a range of analytical techniques to evaluate their phase composition, structural properties, chemical composition, thickness, and surface morphology. Grazing-incidence X-ray diffraction (GI-XRD) was performed with a Philips PW‐1730 X-ray diffractometer, utilizing Cu(Kα) radiation to record diffraction patterns from 2θ = 10° to 80° with a step size of 0.02°. Incidence angles of α = 3° were employed. The surface morphology and microstructural features were examined through field emission scanning electron microscopy (FE-SEM), while elemental mapping was performed using energy-dispersive X-ray spectroscopy (EDS). Additionally, atomic force microscopy (AFM, NanoscopeVMulti Mode System, Veeco) was utilized to assess surface roughness and topographical characteristics. To investigate the electrochemical properties of the synthesized samples, a three-electrode cell was used containing a platinum mesh as a counter electrode, a Ag/AgCl as a reference electrode, and the samples as a working electrode. The corrosion behavior of the deposited coatings under different conditions was investigated by Electrochemical impedance spectroscopy (EIS) test using a potentiostat/ galvanostat (Metrohm, Autolab IMP) in SBF solution, which prepared with a specific salt composition [60], for 1 h and 24 h at 37 °C and pH 7.4 with a scan rate of +1 mV/s. In addition, EIS was conducted in the range of 10-1 to 105 Hz by applying 10 mV as the amplitude and a frequency of 20 mHz in the sweeping mode. Potentiodynamic polarization was conducted in the range of -250 mV to +250 mV versus the open circuit potential at a scanning rate of 1 mV.s-1. The corrosion current (icorr) and potential (Ecorr) were determined by Tafel extrapolation [61]. The nano-mechanical properties such as Young's modulus (E) and hardness were obtained by nano-indentation and nano-scratch tests on the HysitronTriboScope. The nano-indenters with a spherical diamond tip (radius of 1 μm) made by D.R. U.D. Hangen, SURFACE, Huckelhoven were installed on the NanoScope III AFM. This test was conducted with the load control method, in which the applied load to the indenter was increased gradually up to a predetermined maximum load and then reduced continuously to zero. The load and relevant diamond tip displacement were recorded and the hardness and elastic modulus were determined from the continuous loading and unloading curves. To improve the statistics, each test was repeated 5 times.
3.1. Phase and structural properties
The GI-XRD patterns of anodized Ti6Al4V, hydroxyapatite (HA) coatings, and HA-collagen composite coatings applied to the anodized Ti6Al4V substrate are presented in Figure. 1. These patterns reveal the intricate interplay of titanium dioxide polymorphs, namely anatase and rutile. This mixed-phase oxide layer is characteristic of titanium anodization processes, which result from the complex interactions of electrochemical reactions and oxide growth dynamics at the titanium-electrolyte interface. The presence of anatase was confirmed by reflections corresponding to the (112), (105), (220), and (215) crystallographic planes, while the rutile phase was identified by its reflections from the (111) and (310) planes. Although the relative intensities of these peaks were not quantified, they qualitatively suggest the distribution of the two phases [53-55].
Figure 1: GI-XRD patterns of anodized Ti6Al4V, hydroxyapatite (HA) coatings, and HA-collagen composite coatings applied to the anodized Ti6Al4V.
Notably, the anatase-to-rutile ratio plays a pivotal role in defining the functional attributes of the oxide layer, particularly its photocatalytic behavior and corrosion resistance. The subsequent deposition of the HA coating introduced new reflections into the GI-XRD patterns, corresponding to the (002), (211), (112), and (300) planes of HA. These peaks, consistent with the ICDD PDF card # 00-009-0432, confirm the successful formation of the HA phase on the anodized titanium substrate. Interestingly, the peaks associated with the underlying anatase and rutile phases remained visible, indicating that the HA coating is not entirely dense. This observation suggests that the coating possesses a degree of porosity, allowing X-rays to penetrate through the HA layer and interact with the underlying oxide. Such porosity is often characteristic of HA coatings produced by various deposition techniques and has both beneficial and detrimental implications. While it provides a favorable structure for cell infiltration and osseointegration, it may also compromise the coating’s mechanical integrity and corrosion resistance. Furthermore, the incorporation of collagen into the HA coating induced subtle yet significant changes in the GI-XRD patterns. While the characteristic peaks of HA retained their positions, their intensities increased noticeably compared to the HA-only coated sample. Concurrently, the intensities of the anatase and rutile substrate peaks were attenuated. These observations suggest that the HA-collagen composite coating is thicker than its HA-only counterpart. It is plausible that collagen, introduced during the deposition process, acted as a growth promoter for the HA phase by influencing its nucleation and growth kinetics. This effect may arise from collagen’s ability to provide nucleation sites or alter the local chemical environment, thereby enhancing crystal growth in SBF solution. magnification applied to the anodized Ti6Al4V substrate.Collagen, being an organic macromolecule, typically does not produce sharp, distinct peaks in XRD patterns. Instead, its presence may manifest as an increase in background noise or a broad, low-intensity feature. While these subtle contributions are challenging to deconvolute without advanced analytical techniques, the overall changes in the XRD patterns provide valuable insights into the structure and composition of the HA-collagen composite coating [53-54].
Figure 2. FESEM images of: (a) anodized Ti6Al4V; (b) hydroxyapatite (HA) coating, (c) HA-collagen coating at 30,000× magnification, ; and (d) HA-collagen coatings at 80,000×.
Figure 3: EDS analyses of: (a) anodized Ti6Al4V substrate; (b) hydroxyapatite (HA) coating; and (c) HA-collagen composite coating applied to the anodized Ti6Al4V substrate
Energy-dispersive X-ray (EDX) analysis (Figure. 3) determined the Ca/P molar ratio of the HA coating to be 1.76, which is close to the ideal stoichiometric ratio of HA (1.67) [62-63]. Minor deviations could stem from surface contamination, slight compositional variations, or the inherent limitations of the EDX technique, which analyzes a relatively large area. The analysis confirmed the presence of key elements (Ca, P, O, Ti, Al, and V) in all three samples. And also, elemental mapping (Figure. 4) provided further insight into the distribution of elements within the coatings. The maps showed a uniform distribution of calcium, phosphorus, and oxygen in both HA and HA-collagen coatings, indicating a controlled and consistent deposition process. For HA-collagen coatings, the nitrogen map revealed an even distribution of collagen throughout the HA matrix, critical for creating a biomimetic environment that supports uniform cell interactions and predictable tissue regeneration.By combining data from SEM, EDX, and elemental mapping, a detailed understanding of the coating structure and composition emerges. SEM highlights how collagen affects HA morphology, creating a thicker coating with aligned needle-like structures. EDX confirms the near-stoichiometric HA composition, while elemental mapping underscores the homogenous dispersion of collagen in the composite. These findings demonstrate the potential of HA-collagen coatings to enhance biological and mechanical properties, making them ideal for biomedical applications.
3.2. Thermodynamic Considerations and Reaction Sequence in Hydroxyapatite Coating Electrodeposition
The extent of surface coverage by these deposits is primarily determined by the nucleation and growth ability of the target material from a thermodynamic perspective. Given this information, it is crucial to carefully control the formation mechanism of the hydroxyapatite layer during the electrodeposition process. Hydroxyapatite coatings are produced through a series of reactions that are essential for successful electrodeposition, they are [64-65]:
2H2O + 2e− → H2 ↑ +2OH− (1)
H2P + OH− → HP
+ H2O (2)
HP + OH− → P
+ H2O (3)
Ca2+ + HP + 2H2O → CaHPO4.2H2O ↓ (4)
5Ca2+ + 3P + OH− → Ca5(PO4)3(OH) ↓ (5)
10CaHPO4 + 2OH− → Ca10(PO4)6(OH)2 ↓ +4P− + 10H+ (6)
The hydroxyl groups (OH−) required for HA synthesis are provided from the water electrolysis reaction according to Eg. 1. Increasing the pH values near the cathode surface promotes the H2P (as the stable phosphate salt at pH value 4.5) to react with hydroxyl ions to create H2P
(see Eq. 2) [64]. The increase in hydroxyl concentration results in an increase in
ions according to Eq. 3. The brushite (dicalcium phosphate dihydrate, DCPD) will be produced as a result of the combination of Ca2+ and HP
ions according to Eq. 4. In addition, hydroxyapatite deposit materials can emerge by the reaction of Ca2+, P
, and hydroxyl ion used on Eq. 5. While alkali treatment, the brushite deposited transforms to HA according to Eq. 6. It must be mentioned that hydrogen peroxide added to the electrolyte provides a supplementary source for OH− generation according to Eq. 4. In such a case, the amount of released hydrogen gas at the cathode surface (Eq. 1) will decrease, which can result in the emergence of coatings having better uniformity and higher adhesion. Nevertheless, it has been proclaimed that the rate of titanium oxidation will increase at the proximity of the H2O2 compound, Eq. 7 [65].
H2O2 + 2e− → 2OH− (7)
Complete surface coverage indicates the great capability of entire surface points to produce sediment from a thermodynamic point of view. It was claimed by Nielsen that the thermodynamic driving force for calcium-phosphate (CaP) precipitation can be measured by Eq. 8 [66]:
ΔG = −RT ln (IAP/KSP)/n (8)
Where T represents the absolute temperature, R is the ideal gas constant, and n shows the number of ions in the precipitate compound material. IAP and KSP represent the ion activity and the thermodynamic solubility product, respectively, that can be granted for HA and DCPD by Eqs. (9-12):
IAP (HA) = (Ca2+)5(P)3(OH−) (9)
IAP (DCPD) = (Ca2+) (HP) (10)
Log KSP (HA) = −1.6657 − 0.098215T − 8219.41/T (11)
Log KSP (DCPD) = 18.180752 − 0.0420307T − 3649.5701/T (12)
In general, it has been taken into account that the precipitation of CaP coatings while electro-deposition mostly relies on the number of hydroxyl ions (see Eq. 2). It has been proved that despite minor variations in the pH values of the bulk solution, the hydroxyl ions concentration will increase considerably around the cathode surface while electro-deposition. In this regard, CaP precipitates are mostly created on or in the vicinity of the Ti6Al4V cathode surface, the local pH values could be a prime dominant factor affecting the creation of calcium phosphate deposits. According to Eqs. 7–9, at higher pH values the HA is the optimal calcium phosphate phase to precipitate from the solution. Based on Faraday’s law in electrolysis, minding that the electricity is entirely consumed bywater reduction and hydroxyl ions production, the maximum theoretical concentration of OH− in the vicinity of the cathode (alayer with the thickness of δ), the surface can be described by Eq. 13 [67]:
[OH−] = It/zFAδ (13)
Where I indicate the electrical current, z indicates the number of transferred electrons, F indicates the Faraday constant, A indicates the surface area, and t represents for the deposition time. In this work, electrodeposition of HA was carried out on different surfaces under similar process parameters (voltage of 3.1 V, deposition time of 15 min, and the same chemical composition of electrolyte). Generally, it could be said that the difference in the concentration of hydroxyl ions in the proximity of the surfaces and as a result the formation of coatings with unique characteristics was exclusively because of the difference in the surface characteristics of the substrate. It is worth mentioning that the oxide layer created on many surfaces of Ti6Al4V alloy will decrease the effective electrical current during the electro-deposition process. With more increase in the thickness of the oxide layer on the surface, the greater the reduction of the hydroxyl concentration according to Eq.13. The lower concentration of hydroxyl ions at the cathode surface will decrease the ability of calcium-phosphate deposits to generate from a thermodynamic point of view. However, it could be observed that the process parameters employed in this work have resulted in almost the entire coverage of the sample surface. Whereas the HAP electrodeposition at V = 3.1 V and deposition time of 15 min on beta-annealed sample resulted in insufficient surface coverage, as can be seen in Fig. 2b.
The Ca/P molar ratios for synthesized coatings were almost the same as that of HA indicating the presence of sufficient amounts of hydroxyl anions to create hydroxyapatite, based on Eqs. 1-6. Considering that along with almost the entire coverage of the surface by deposits, it can be confirmed that the deposition voltage employed in this work was enough to compensate for the effective current reduction because of the presence of an oxide layer, and as a result, by creating enough amount of hydroxyl ions the coatings would assumingly comprise hydroxyapatite as the prime compound. It can be concluded that the Ca/P molar ratios for calcium phosphate compounds (excluding HA) may have emerged while the electrodeposition process changed from 0.5 for calcium dihydrogen orthophosphate (Ca(HPO4)2) to 1.5 for three calcium phosphate (Ca3(PO4)2) [62-63]. As a result, the presence of such compounds in the structure of the coating layer causes the deviation of the Ca/P ratio from the Ca/P molar ratio of HAP (i.e. 1.76). Taking into account, the Ca/P ratio of the coatings formed in this work could indicate the creation of a higher amount of hydroxyapatite compared to other calcium phosphate compounds during the electro-deposition process.
However, the nucleation and growth mechanism of Ca-P coatings on Ti substrate change while deposition and cause various morphologies at different stages of deposition. Studies [62-67], showed that at the first stage of deposition (t = 1 min), the electrolyte is highly supersaturated and the deposited coatings are polycrystalline HA. The morphology of the coatings at this stage is randomly oriented and highly branched nanoplates. At the second stage (t = 3 min), crystal growth progresses along the b- and c-axes and forms micro-sized plates. At the third stage of deposition, the degree of supersaturation is reduced (t > 10 min), and the deposited crystals propagate along the c-axis and become plate-like or needle-like crystals. According to the above and the electrodeposition time employed in this work (i.e. 15 min), the length of the largest dimension of the plate-likes and needle-likedeposits was assumed as a measure of HAP deposit sizes.
Indeed, plate-like and needle-like HAP crystals in the microstructure (Figure. 2b, 2c and 2d) show that the nucleation and growth of HA crystals are entirely performed on different surfaces. After the nucleation stage, HA will initially generate with spherical crystals. As the deposition time increases, spherical HA crystals will agglomerate to create plate-like crystals. Needle-like morphology is created due to the growth of HA crystals outwards of the plate-like crystals [68- 69].
3.3. Nanomechanical Characterization
Fig. 5 presents AFM images, capturing the surface morphology of the synthesized HA coating and HA- Collagen nanostructured coating deposited on the anodized Ti-6Al-4V substrate within a 5×5 µm² area. These images offer valuable insights into the surface roughness and the arrangement of collagen fibers within the composite structure. The presence of collagen fibers on the surface is evident, with a predominantly parallel orientation. However, discerning individual fibers proves challenging due to the partial obscuring of inter-fiber boundaries. This observation suggests a close packing or intertwining of the collagen fibrils, creating a relatively smooth surface texture at this scale. Actually, comparing the AFM images of the HA and HA-collagen coatings (Figs. 5a- 5b), a noticeable reduction in surface roughness is observed upon collagen incorporation. This reduction suggests that collagen effectively fills some of the pores and irregularities present in the HA structure, leading to a smoother surface finish. This smoothing effect can have profound implications for cell-surface interactions, potentially enhancing cell adhesion, spreading, and proliferation. Furthermore, the presence of collagen can impart a degree of flexibility and resilience to the coating, potentially influencing its mechanical behavior.
Figure 5: AFM image of: (a) hydroxyapatite (HA) coating; and (b) HA-collagen coating applied to the anodized Ti6Al4V substrate.
The nanomechanical properties of the coatings were investigated using nanoindentation and nanoscratch techniques, performed with an AFM-based nanoindenter. The nanoindentation results, presented in Fig. 6 and summarized in Table 1, reveal a clear trend in both hardness and Young's modulus. As seen in Table 1, both hardness (H) and Young’s modulus (E) exhibit a decreasing trend from the anodized Ti substrate to the HA coating and further to the HA-collagen nanostructured coating. Specifically, the Young's modulus decreases from 0.4 GPa for anodized Ti to 0.2 GPa for the HA coating and subsequently to 0.1 GPa for the HA-collagen composite. A similar trend is observed for hardness, decreasing from 0.18 GPa for anodized Ti to 0.07 GPa for the HA coating and finally to 0.01 GPa for the HA-collagen composite. This reduction in both hardness and modulus with the incorporation of HA and, more significantly, with the addition of collagen, is not entirely unexpected. HA, while inherently hard, can exhibit reduced mechanical properties in coatings due to its often porous microstructure. The introduction of collagen, a soft and flexible biopolymer, further reduces the overall stiffness and hardness of the composite material. This decrease in stiffness and hardness can be advantageous in certain biomedical applications, particularly for bone implants, as it can bring the mechanical properties of the coating closer to that of natural bone, mitigating stress shielding effects.
Figure 6: Nano-indentation test of: (a) anodized Ti6Al4V substrate; (b) hydroxyapatite (HA) coating; and (c) HA-collagen composite coating applied to the anodized Ti6Al4V substrate.
Table 1: Nanoindentation test data for anodized Ti6Al4V, hydroxyapatite (HA) coating and HA-Collagen nanostructured coating applied on anodized Ti6Al4V
And also, the tribological behavior of the coatings was assessed using nanoscratch testing, the results of which are depicted in Figure. 7. The coefficient of friction (COF) for the anodized titanium shows an increase from an initial value of 0.61 to 0.92 as the scratch test progresses. A similar trend is observed for the HA coating, with the COF increasing from 0.62 to 0.98. This increase in COF during scratching is likely due to wear mechanisms, such as ploughing and micro-cutting, leading to increased surface roughness and frictional resistance. However, the HA-collagen composite exhibits a distinct behavior. The initial COF is higher (1.12), but it decreases to 0.94 as the scratch test proceeds. This initial higher COF could be a consequence of the presence of the relatively soft collagen on the surface, which might initially deform more readily under the indenter
Figure 6: Evolution of the coefficient of friction over time during nano scratch testing of: anodized Ti6Al4V substrate; (b) hydroxyapatite (HA) coating; and (c) HA-collagen composite coating applied to the anodized Ti6Al4V substrate. Test conditions: sliding speed of 1 m/s and applied pressure of 3.45 MPa.
The subsequent decrease in COF suggests that the collagen may be acting as a lubricant, reducing the friction between the indenter and the coating surface. Furthermore, the fibrous nature of collagen could enhance the coating's resistance to ploughing and material removal, preventing a sharp increase in COF typically observed in brittle materials.
The combined analysis of AFM, nanoindentation, and nanoscratch testing provides a holistic understanding of the coating's structure, mechanical properties, and tribological performance. The AFM observations reveal that collagen fills pores in the HA layer, reducing surface roughness. Nanoindentation demonstrates that the incorporation of HA and, especially, collagen reduces both hardness and Young’s modulus, a desirable trait for bone implant applications. Nanoscratch testing suggests that collagen can reduce the COF and potentially improve wear resistance. These interconnected findings suggest that combining HA with collagen creates biomimetic coatings with enhanced properties for biomedical applications. Specifically, the reduction in roughness, the tuning of mechanical properties closer to bone, and the improved tribological behavior can collectively contribute to better cell adhesion, reduced implant wear, and improved biocompatibility.
3.4. Corrosion behavior and its mechanism
The nyquist plots and extracted data acquired for different coated surfaces in 1 h and 24h SBF immersion periods are shown in Fig. 7. The corresponding equivalent circuit for the EIS results is depicted in Table 2 in which Rs, Rp, and Rct are solution resistance, the resistance of the coating, which are directly linked to the coating defects such as porosity and microcracks, and charge transfer resistance of electrical double layer (EDL), respectively. CPEc and CPEdl are constant phase elements for coating and EDL, respectively. The fitting curves (solid lines) of the EIS spectra are incorporated into experimental data (scatter points), which are well consistent with each other.
Referring to the fitting results of EIS (Table. 2), the charge transfer resistances of different surfaces (Rct) have a much higher value compared to the corresponding coating resistance (Rc). This is a rather interesting finding. It suggests that, despite the beneficial effects of the HA coating and, even more so, the HA-collagen composite, the overall corrosion resistance of each surface is primarily governed by the charge transfer resistance at the substrate-oxide interface. In other words, the inherent protective properties of the naturally formed or anodically grown oxide layer on the Ti6Al4V substrate play a dominant role in determining the corrosion rate. The coatings, while offering additional protection, primarily influence the impedance behavior at higher frequencies, related to their own dielectric and barrier properties. The anodized Ti6Al4V has the lowest Rct value of 0.89 MΩ.cm2, which implies a higher corrosion rate than coated samples. The highest Rct value belongs to HA-Collagen nanostructured coating, which means a lower corrosion rate in SBF solution, suggesting that the incorporation of collagen within the HA coating provides a synergistic effect, further enhancing the corrosion protection. This could be attributed to several factors, including a reduction in coating porosity (as observed in the Figure. 2 and Figure. 5), improved barrier properties due to the presence of the organic collagen network, and a possible influence on the electrochemical processes at the interface
Figure 7: The nyquist plot for anodized Ti-6Al-4V, HA coating, and HA- Collagen nanostructured coating after, a) 1 hour and, b) 24 hours times immersion in SBF solution at 37.5±1.5˚C.
Table 2: EIS data of anodized Ti-6Al-4V, HA coating, and HA- Collagen nanostructured coating after 1 hour and 24 hours times immersion in SBF solution at 37.5±1.5˚C.
Comparatively, the HA- Collagen nanostructured coating has excellent corrosion resistance (Table 2 and Fig.7b) for 24 h immersion in SBF solution compared to its counterparts. As a result, the HA- Collagen nanostructured coating has the highest protective efficiency and in turn outstanding biocompatibility because of the electrochemically deposited HA and particularly sol-gel derived collagen layer on top. These results are in complete agreement with corrosion current densities obtained from polarization curves in Fig. 8, which shows the electrochemical polarization plots for the anodized Ti6Al4V, HA coating and HA- Collagen nanostructured coating, and the results summarized in Table 3. Results indicate that the corrosion rate of anodized Ti6Al4V decreases considerably unlike the sample having optimum coating HA-Collagen layers on its surface. The extremely low corrosion rate is critical in implant applications since it results in the negligible release of toxic metallic ions to the surrounding tissues [17-19].
Figure 8: Polarization plot for anodized Ti-6Al-4V, HA coating, and HA-Collagen nanostructured coating in SBF solution at 37.5±1.5˚C.
Table 3: Summerized data of polarization test of anodized Ti-6Al-4V, HA coating, and HA-Collagen nanostructured coating in SBF solution at 37.5±1.5˚C.
Polarization resistance (Rp) values are calculated according to the Eq. (14) [67], and show in Fig. 9:
Rp = (βa × βc) / (2.303 × icorr × (βa + βc)) (14)
where, icorrr indicates the corrosion current density, βa and βc are the anodic and cathodic Tafel constants, respectively, and results show a significant increase in the polarization resistance of the coating containing HA and collagen, so that the polarization resistance has increased by more than 600% compared to anodized titanium and also HA coating. Finally, the observed enhancement in corrosion resistance with the HA-collagen coating suggests its potential for improving the long-term performance and biocompatibility of Ti6Al4V implants.
Figure 9: Polarization resistance (Rp) values for anodized Ti-6Al-4V, HA coating, and HA- Collagen nanostructured coating.
This study meticulously investigated the synergistic effects of combining hydroxyapatite (HA) with collagen on anodized Ti-6Al-4V, exploring the resulting coatings' structural, mechanical, tribological, and corrosion properties. GI-XRD analysis confirms the successful development of a multi-layered coating on Ti6Al4V, comprising an anodized titanium layer with mixed anatase-rutile structure, followed by a crystalline HA coating. The addition of collagen enhances HA peak intensities, indicating a thicker, more integrated composite layer. Nanoindentation tests reveal decreasing hardness and Young's modulus from anodized Ti to HA, and further to HA-collagen coatings. This softening, mainly due to collagen addition, potentially benefits bone implants by mimicking natural bone properties. Nanoscratch tests show increasing coefficient of friction (COF) for anodized Ti and HA coatings during scratching, likely due to wear mechanisms. Interestingly, the HA-collagen composite exhibits a unique behavior with an initially higher COF that decreases during testing, suggesting improved tribological properties. EIS and polarization tests reveal that the HA-Collagen nanostructured coating significantly enhances Ti6Al4V implant corrosion resistance in SBF. This composite coating shows a 600% increase in polarization resistance compared to anodized titanium and HA coating alone, exhibiting the highest charge transfer resistance (0.89 MΩ.cm²) and lowest corrosion rate. The observed reduction in roughness, the tailored mechanical properties, the improved wear resistance, and the enhanced corrosion protection collectively contribute to a more biocompatible and durable coating. These findings strongly suggest that the HA-collagen composite holds considerable promise for enhancing the long-term performance and clinical success of biomedical implants, particularly in orthopedic and dental applications.
The work was financially supported by Malayer University Research Grant, No. 1400.
Clearly Auctoresonline and particularly Psychology and Mental Health Care Journal is dedicated to improving health care services for individuals and populations. The editorial boards' ability to efficiently recognize and share the global importance of health literacy with a variety of stakeholders. Auctoresonline publishing platform can be used to facilitate of optimal client-based services and should be added to health care professionals' repertoire of evidence-based health care resources.
Journal of Clinical Cardiology and Cardiovascular Intervention The submission and review process was adequate. However I think that the publication total value should have been enlightened in early fases. Thank you for all.
Journal of Women Health Care and Issues By the present mail, I want to say thank to you and tour colleagues for facilitating my published article. Specially thank you for the peer review process, support from the editorial office. I appreciate positively the quality of your journal.
Journal of Clinical Research and Reports I would be very delighted to submit my testimonial regarding the reviewer board and the editorial office. The reviewer board were accurate and helpful regarding any modifications for my manuscript. And the editorial office were very helpful and supportive in contacting and monitoring with any update and offering help. It was my pleasure to contribute with your promising Journal and I am looking forward for more collaboration.
We would like to thank the Journal of Thoracic Disease and Cardiothoracic Surgery because of the services they provided us for our articles. The peer-review process was done in a very excellent time manner, and the opinions of the reviewers helped us to improve our manuscript further. The editorial office had an outstanding correspondence with us and guided us in many ways. During a hard time of the pandemic that is affecting every one of us tremendously, the editorial office helped us make everything easier for publishing scientific work. Hope for a more scientific relationship with your Journal.
The peer-review process which consisted high quality queries on the paper. I did answer six reviewers’ questions and comments before the paper was accepted. The support from the editorial office is excellent.
Journal of Neuroscience and Neurological Surgery. I had the experience of publishing a research article recently. The whole process was simple from submission to publication. The reviewers made specific and valuable recommendations and corrections that improved the quality of my publication. I strongly recommend this Journal.
Dr. Katarzyna Byczkowska My testimonial covering: "The peer review process is quick and effective. The support from the editorial office is very professional and friendly. Quality of the Clinical Cardiology and Cardiovascular Interventions is scientific and publishes ground-breaking research on cardiology that is useful for other professionals in the field.
Thank you most sincerely, with regard to the support you have given in relation to the reviewing process and the processing of my article entitled "Large Cell Neuroendocrine Carcinoma of The Prostate Gland: A Review and Update" for publication in your esteemed Journal, Journal of Cancer Research and Cellular Therapeutics". The editorial team has been very supportive.
Testimony of Journal of Clinical Otorhinolaryngology: work with your Reviews has been a educational and constructive experience. The editorial office were very helpful and supportive. It was a pleasure to contribute to your Journal.
Dr. Bernard Terkimbi Utoo, I am happy to publish my scientific work in Journal of Women Health Care and Issues (JWHCI). The manuscript submission was seamless and peer review process was top notch. I was amazed that 4 reviewers worked on the manuscript which made it a highly technical, standard and excellent quality paper. I appreciate the format and consideration for the APC as well as the speed of publication. It is my pleasure to continue with this scientific relationship with the esteem JWHCI.
This is an acknowledgment for peer reviewers, editorial board of Journal of Clinical Research and Reports. They show a lot of consideration for us as publishers for our research article “Evaluation of the different factors associated with side effects of COVID-19 vaccination on medical students, Mutah university, Al-Karak, Jordan”, in a very professional and easy way. This journal is one of outstanding medical journal.
Dear Hao Jiang, to Journal of Nutrition and Food Processing We greatly appreciate the efficient, professional and rapid processing of our paper by your team. If there is anything else we should do, please do not hesitate to let us know. On behalf of my co-authors, we would like to express our great appreciation to editor and reviewers.
As an author who has recently published in the journal "Brain and Neurological Disorders". I am delighted to provide a testimonial on the peer review process, editorial office support, and the overall quality of the journal. The peer review process at Brain and Neurological Disorders is rigorous and meticulous, ensuring that only high-quality, evidence-based research is published. The reviewers are experts in their fields, and their comments and suggestions were constructive and helped improve the quality of my manuscript. The review process was timely and efficient, with clear communication from the editorial office at each stage. The support from the editorial office was exceptional throughout the entire process. The editorial staff was responsive, professional, and always willing to help. They provided valuable guidance on formatting, structure, and ethical considerations, making the submission process seamless. Moreover, they kept me informed about the status of my manuscript and provided timely updates, which made the process less stressful. The journal Brain and Neurological Disorders is of the highest quality, with a strong focus on publishing cutting-edge research in the field of neurology. The articles published in this journal are well-researched, rigorously peer-reviewed, and written by experts in the field. The journal maintains high standards, ensuring that readers are provided with the most up-to-date and reliable information on brain and neurological disorders. In conclusion, I had a wonderful experience publishing in Brain and Neurological Disorders. The peer review process was thorough, the editorial office provided exceptional support, and the journal's quality is second to none. I would highly recommend this journal to any researcher working in the field of neurology and brain disorders.
Dear Agrippa Hilda, Journal of Neuroscience and Neurological Surgery, Editorial Coordinator, I trust this message finds you well. I want to extend my appreciation for considering my article for publication in your esteemed journal. I am pleased to provide a testimonial regarding the peer review process and the support received from your editorial office. The peer review process for my paper was carried out in a highly professional and thorough manner. The feedback and comments provided by the authors were constructive and very useful in improving the quality of the manuscript. This rigorous assessment process undoubtedly contributes to the high standards maintained by your journal.
International Journal of Clinical Case Reports and Reviews. I strongly recommend to consider submitting your work to this high-quality journal. The support and availability of the Editorial staff is outstanding and the review process was both efficient and rigorous.
Thank you very much for publishing my Research Article titled “Comparing Treatment Outcome Of Allergic Rhinitis Patients After Using Fluticasone Nasal Spray And Nasal Douching" in the Journal of Clinical Otorhinolaryngology. As Medical Professionals we are immensely benefited from study of various informative Articles and Papers published in this high quality Journal. I look forward to enriching my knowledge by regular study of the Journal and contribute my future work in the field of ENT through the Journal for use by the medical fraternity. The support from the Editorial office was excellent and very prompt. I also welcome the comments received from the readers of my Research Article.
Dear Erica Kelsey, Editorial Coordinator of Cancer Research and Cellular Therapeutics Our team is very satisfied with the processing of our paper by your journal. That was fast, efficient, rigorous, but without unnecessary complications. We appreciated the very short time between the submission of the paper and its publication on line on your site.
I am very glad to say that the peer review process is very successful and fast and support from the Editorial Office. Therefore, I would like to continue our scientific relationship for a long time. And I especially thank you for your kindly attention towards my article. Have a good day!
"We recently published an article entitled “Influence of beta-Cyclodextrins upon the Degradation of Carbofuran Derivatives under Alkaline Conditions" in the Journal of “Pesticides and Biofertilizers” to show that the cyclodextrins protect the carbamates increasing their half-life time in the presence of basic conditions This will be very helpful to understand carbofuran behaviour in the analytical, agro-environmental and food areas. We greatly appreciated the interaction with the editor and the editorial team; we were particularly well accompanied during the course of the revision process, since all various steps towards publication were short and without delay".
I would like to express my gratitude towards you process of article review and submission. I found this to be very fair and expedient. Your follow up has been excellent. I have many publications in national and international journal and your process has been one of the best so far. Keep up the great work.
We are grateful for this opportunity to provide a glowing recommendation to the Journal of Psychiatry and Psychotherapy. We found that the editorial team were very supportive, helpful, kept us abreast of timelines and over all very professional in nature. The peer review process was rigorous, efficient and constructive that really enhanced our article submission. The experience with this journal remains one of our best ever and we look forward to providing future submissions in the near future.
I am very pleased to serve as EBM of the journal, I hope many years of my experience in stem cells can help the journal from one way or another. As we know, stem cells hold great potential for regenerative medicine, which are mostly used to promote the repair response of diseased, dysfunctional or injured tissue using stem cells or their derivatives. I think Stem Cell Research and Therapeutics International is a great platform to publish and share the understanding towards the biology and translational or clinical application of stem cells.
I would like to give my testimony in the support I have got by the peer review process and to support the editorial office where they were of asset to support young author like me to be encouraged to publish their work in your respected journal and globalize and share knowledge across the globe. I really give my great gratitude to your journal and the peer review including the editorial office.
I am delighted to publish our manuscript entitled "A Perspective on Cocaine Induced Stroke - Its Mechanisms and Management" in the Journal of Neuroscience and Neurological Surgery. The peer review process, support from the editorial office, and quality of the journal are excellent. The manuscripts published are of high quality and of excellent scientific value. I recommend this journal very much to colleagues.
Dr.Tania Muñoz, My experience as researcher and author of a review article in The Journal Clinical Cardiology and Interventions has been very enriching and stimulating. The editorial team is excellent, performs its work with absolute responsibility and delivery. They are proactive, dynamic and receptive to all proposals. Supporting at all times the vast universe of authors who choose them as an option for publication. The team of review specialists, members of the editorial board, are brilliant professionals, with remarkable performance in medical research and scientific methodology. Together they form a frontline team that consolidates the JCCI as a magnificent option for the publication and review of high-level medical articles and broad collective interest. I am honored to be able to share my review article and open to receive all your comments.
“The peer review process of JPMHC is quick and effective. Authors are benefited by good and professional reviewers with huge experience in the field of psychology and mental health. The support from the editorial office is very professional. People to contact to are friendly and happy to help and assist any query authors might have. Quality of the Journal is scientific and publishes ground-breaking research on mental health that is useful for other professionals in the field”.
Dear editorial department: On behalf of our team, I hereby certify the reliability and superiority of the International Journal of Clinical Case Reports and Reviews in the peer review process, editorial support, and journal quality. Firstly, the peer review process of the International Journal of Clinical Case Reports and Reviews is rigorous, fair, transparent, fast, and of high quality. The editorial department invites experts from relevant fields as anonymous reviewers to review all submitted manuscripts. These experts have rich academic backgrounds and experience, and can accurately evaluate the academic quality, originality, and suitability of manuscripts. The editorial department is committed to ensuring the rigor of the peer review process, while also making every effort to ensure a fast review cycle to meet the needs of authors and the academic community. Secondly, the editorial team of the International Journal of Clinical Case Reports and Reviews is composed of a group of senior scholars and professionals with rich experience and professional knowledge in related fields. The editorial department is committed to assisting authors in improving their manuscripts, ensuring their academic accuracy, clarity, and completeness. Editors actively collaborate with authors, providing useful suggestions and feedback to promote the improvement and development of the manuscript. We believe that the support of the editorial department is one of the key factors in ensuring the quality of the journal. Finally, the International Journal of Clinical Case Reports and Reviews is renowned for its high- quality articles and strict academic standards. The editorial department is committed to publishing innovative and academically valuable research results to promote the development and progress of related fields. The International Journal of Clinical Case Reports and Reviews is reasonably priced and ensures excellent service and quality ratio, allowing authors to obtain high-level academic publishing opportunities in an affordable manner. I hereby solemnly declare that the International Journal of Clinical Case Reports and Reviews has a high level of credibility and superiority in terms of peer review process, editorial support, reasonable fees, and journal quality. Sincerely, Rui Tao.
Clinical Cardiology and Cardiovascular Interventions I testity the covering of the peer review process, support from the editorial office, and quality of the journal.
Clinical Cardiology and Cardiovascular Interventions, we deeply appreciate the interest shown in our work and its publication. It has been a true pleasure to collaborate with you. The peer review process, as well as the support provided by the editorial office, have been exceptional, and the quality of the journal is very high, which was a determining factor in our decision to publish with you.
The peer reviewers process is quick and effective, the supports from editorial office is excellent, the quality of journal is high. I would like to collabroate with Internatioanl journal of Clinical Case Reports and Reviews journal clinically in the future time.
Clinical Cardiology and Cardiovascular Interventions, I would like to express my sincerest gratitude for the trust placed in our team for the publication in your journal. It has been a true pleasure to collaborate with you on this project. I am pleased to inform you that both the peer review process and the attention from the editorial coordination have been excellent. Your team has worked with dedication and professionalism to ensure that your publication meets the highest standards of quality. We are confident that this collaboration will result in mutual success, and we are eager to see the fruits of this shared effort.
Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, I hope this message finds you well. I want to express my utmost gratitude for your excellent work and for the dedication and speed in the publication process of my article titled "Navigating Innovation: Qualitative Insights on Using Technology for Health Education in Acute Coronary Syndrome Patients." I am very satisfied with the peer review process, the support from the editorial office, and the quality of the journal. I hope we can maintain our scientific relationship in the long term.
Dear Monica Gissare, - Editorial Coordinator of Nutrition and Food Processing. ¨My testimony with you is truly professional, with a positive response regarding the follow-up of the article and its review, you took into account my qualities and the importance of the topic¨.
Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, The review process for the article “The Handling of Anti-aggregants and Anticoagulants in the Oncologic Heart Patient Submitted to Surgery” was extremely rigorous and detailed. From the initial submission to the final acceptance, the editorial team at the “Journal of Clinical Cardiology and Cardiovascular Interventions” demonstrated a high level of professionalism and dedication. The reviewers provided constructive and detailed feedback, which was essential for improving the quality of our work. Communication was always clear and efficient, ensuring that all our questions were promptly addressed. The quality of the “Journal of Clinical Cardiology and Cardiovascular Interventions” is undeniable. It is a peer-reviewed, open-access publication dedicated exclusively to disseminating high-quality research in the field of clinical cardiology and cardiovascular interventions. The journal's impact factor is currently under evaluation, and it is indexed in reputable databases, which further reinforces its credibility and relevance in the scientific field. I highly recommend this journal to researchers looking for a reputable platform to publish their studies.
Dear Editorial Coordinator of the Journal of Nutrition and Food Processing! "I would like to thank the Journal of Nutrition and Food Processing for including and publishing my article. The peer review process was very quick, movement and precise. The Editorial Board has done an extremely conscientious job with much help, valuable comments and advices. I find the journal very valuable from a professional point of view, thank you very much for allowing me to be part of it and I would like to participate in the future!”
Dealing with The Journal of Neurology and Neurological Surgery was very smooth and comprehensive. The office staff took time to address my needs and the response from editors and the office was prompt and fair. I certainly hope to publish with this journal again.Their professionalism is apparent and more than satisfactory. Susan Weiner
My Testimonial Covering as fellowing: Lin-Show Chin. The peer reviewers process is quick and effective, the supports from editorial office is excellent, the quality of journal is high. I would like to collabroate with Internatioanl journal of Clinical Case Reports and Reviews.
My experience publishing in Psychology and Mental Health Care was exceptional. The peer review process was rigorous and constructive, with reviewers providing valuable insights that helped enhance the quality of our work. The editorial team was highly supportive and responsive, making the submission process smooth and efficient. The journal's commitment to high standards and academic rigor makes it a respected platform for quality research. I am grateful for the opportunity to publish in such a reputable journal.
My experience publishing in International Journal of Clinical Case Reports and Reviews was exceptional. I Come forth to Provide a Testimonial Covering the Peer Review Process and the editorial office for the Professional and Impartial Evaluation of the Manuscript.
I would like to offer my testimony in the support. I have received through the peer review process and support the editorial office where they are to support young authors like me, encourage them to publish their work in your esteemed journals, and globalize and share knowledge globally. I really appreciate your journal, peer review, and editorial office.
Dear Agrippa Hilda- Editorial Coordinator of Journal of Neuroscience and Neurological Surgery, "The peer review process was very quick and of high quality, which can also be seen in the articles in the journal. The collaboration with the editorial office was very good."
I would like to express my sincere gratitude for the support and efficiency provided by the editorial office throughout the publication process of my article, “Delayed Vulvar Metastases from Rectal Carcinoma: A Case Report.” I greatly appreciate the assistance and guidance I received from your team, which made the entire process smooth and efficient. The peer review process was thorough and constructive, contributing to the overall quality of the final article. I am very grateful for the high level of professionalism and commitment shown by the editorial staff, and I look forward to maintaining a long-term collaboration with the International Journal of Clinical Case Reports and Reviews.
To Dear Erin Aust, I would like to express my heartfelt appreciation for the opportunity to have my work published in this esteemed journal. The entire publication process was smooth and well-organized, and I am extremely satisfied with the final result. The Editorial Team demonstrated the utmost professionalism, providing prompt and insightful feedback throughout the review process. Their clear communication and constructive suggestions were invaluable in enhancing my manuscript, and their meticulous attention to detail and dedication to quality are truly commendable. Additionally, the support from the Editorial Office was exceptional. From the initial submission to the final publication, I was guided through every step of the process with great care and professionalism. The team's responsiveness and assistance made the entire experience both easy and stress-free. I am also deeply impressed by the quality and reputation of the journal. It is an honor to have my research featured in such a respected publication, and I am confident that it will make a meaningful contribution to the field.
"I am grateful for the opportunity of contributing to [International Journal of Clinical Case Reports and Reviews] and for the rigorous review process that enhances the quality of research published in your esteemed journal. I sincerely appreciate the time and effort of your team who have dedicatedly helped me in improvising changes and modifying my manuscript. The insightful comments and constructive feedback provided have been invaluable in refining and strengthening my work".
I thank the ‘Journal of Clinical Research and Reports’ for accepting this article for publication. This is a rigorously peer reviewed journal which is on all major global scientific data bases. I note the review process was prompt, thorough and professionally critical. It gave us an insight into a number of important scientific/statistical issues. The review prompted us to review the relevant literature again and look at the limitations of the study. The peer reviewers were open, clear in the instructions and the editorial team was very prompt in their communication. This journal certainly publishes quality research articles. I would recommend the journal for any future publications.
Dear Jessica Magne, with gratitude for the joint work. Fast process of receiving and processing the submitted scientific materials in “Clinical Cardiology and Cardiovascular Interventions”. High level of competence of the editors with clear and correct recommendations and ideas for enriching the article.
We found the peer review process quick and positive in its input. The support from the editorial officer has been very agile, always with the intention of improving the article and taking into account our subsequent corrections.
My article, titled 'No Way Out of the Smartphone Epidemic Without Considering the Insights of Brain Research,' has been republished in the International Journal of Clinical Case Reports and Reviews. The review process was seamless and professional, with the editors being both friendly and supportive. I am deeply grateful for their efforts.
To Dear Erin Aust – Editorial Coordinator of Journal of General Medicine and Clinical Practice! I declare that I am absolutely satisfied with your work carried out with great competence in following the manuscript during the various stages from its receipt, during the revision process to the final acceptance for publication. Thank Prof. Elvira Farina